Siwenoid uses a Lantronix UDS1100 server to communicate with the Siemens Algorex CS11 system. For this, the UDS1100 must be updated to a special firmware which is provided by our company. After the update Siwenoid can connect to an UDS1100 through LAN or internet. The serial port of the UDS1100 is connected to the E3H020/21 gateway card of the CS11 system. Gateway communication has to be set to 1200 baud and ISO1745 protocol. A CS11 system can interface with several control units.
UDS1100 parameters can be edited through telnet (9999 port). A firmware specific http://x.x.x.x/index.cgi web server page can be reached in a UDS1100. It shows the application specific status of the communication ports.
Configuration of the UDS1100 can be done from command prompt, via Telnet
The telnet client is part of windows XP, but it needs to be installed in Windows 7. To install telnet use the Windows services menu and install telnet client You can connect to the device server as seen on the picture:
After successfully accessing the device you can see a detailed list with a numbered menu. Press the corresponding number to access the selected menu.
Here the devices LAN configuration can be set up. Set the IP adress, default gateway and the subnet mask.
Here you can setup the rs232 side of the lantronix
After finishing the necessary settings, use menu item 9 (hit 9 on keybopard) to save the settings and exit the configuration utility.
If we did the settings right, then the device should be capable of communicating.